By turning the valve a quarter counter-clockwise it is turned from the
‘talk’ (right figure) to the ‘walk’ (Off=right figure) position (see video
clip). When turning, inside the valve a magnetic pin is rotated towards
the magnet at the end of the membrane. Thereby, the membrane is trapped
and it will not close on exhalation. Explain this feature to the patient
and show the patient how he or she can see in what position the valve
is. It also should be mentioned, that physical exertion with the valve
in talk position can lead to a slightly increased airflow resistance,
even if the valve does not close completely. It is advised to insert the
valve in the housing for speaking adjusted in the talk position. Practice
a couple of times how to adjust the valve when placed in the housing. |